TENMARS TM-195 3-Axis RF Field Strength Meter Instruction Manual

Awoṣe: TM-195

1. Ifihan

The TENMARS TM-195 is a 3-axis high-frequency electromagnetic field strength meter designed for measuring and monitoring Radio Frequency (RF) electromagnetic fields. It is capable of measuring frequencies ranging from 50MHz to 3.5GHz. This manual provides essential information for the safe and effective operation of your device.

2. Alaye Aabo

Please read and understand all safety instructions before operating the device.

  • Má ṣe gbìyànjú láti tún ẹ̀rọ náà ṣe tàbí láti ṣe àtúnṣe rẹ̀. Tọ́ka gbogbo iṣẹ́ ìtọ́jú náà sí àwọn òṣìṣẹ́ tó mọ̀ nípa rẹ̀.
  • Ensure the battery compartment is securely closed during operation.
  • Yago fun ṣiṣafihan ẹrọ naa si awọn iwọn otutu to gaju, ọriniinitutu, tabi oorun taara.
  • Jeki ẹrọ naa kuro ni awọn aaye oofa ti o lagbara.
  • Sọ awọn batiri sọnu ni ibamu si awọn ilana agbegbe.

3. Ọja Ipariview

The TM-195 features a compact design with an integrated 3-axis isotropic sensor for comprehensive RF field strength measurements.

TENMARS TM-195 3-Axis RF Field Strength Meter with carrying case

Nọmba 1: TENMARS TM-195 RF Field Strength Meter and Carrying Case. The meter features an orange spherical sensor at the top, a green LCD display showing '360.4 mV/m' and '0945 XYZ', and control buttons below the display. A black carrying case is visible to the right.

3.1. Awọn irinše

  • 3-Axis Isotropic Sensor: Spherical sensor for comprehensive field detection.
  • Ifihan LCD: 4-1/2 digits, maximum reading 19999, shows measured values, units, and status indicators.
  • Awọn bọtini Iṣakoso: For power, mode selection, unit change, data hold, alarm settings, and data memory.
  • Abala Batiri: Located at the rear for 9V battery.

3.2. Awọn afihan Ifihan

  • XYZ: Indicates triaxial measurement mode.
  • mV/m, V/m, µA/m, mA/m, µW/m², mW/m², µW/cm²: Awọn iwọn wiwọn.
  • Max/AVG: Maximum and Average value indicators.
  • MU: Iṣẹ́ ìdúró dátà ń ṣiṣẹ́.
  • IKỌ: Alarm function active.
  • Aami batiri: Tọkasi ipele batiri.

4. Eto

4.1. Fifi sori batiri

  1. Ṣii ideri iyẹwu batiri ni ẹhin ẹrọ naa.
  2. Insert a new 9V battery (NEDA 1604, IEC 6F22, or JIS 006P), observing polarity.
  3. Pa ideri batiri naa ni aabo.

4.2. Agbara akọkọ Lori

Tẹ awọn Agbara button (labeled with a circle and vertical line) to turn on the device. The display will illuminate and show the current RF field strength.

5. Awọn ilana Iṣiṣẹ

5.1. Agbara Tan / Paa

  • Lati mu agbara ṣiṣẹ, tẹ bọtini naa Agbara bọtini.
  • Lati fi agbara si pipa, tẹ mọlẹ Agbara bọtini fun isunmọ 2 aaya.

5.2. Measurement Modes (MAX/AVG)

Tẹ awọn MAX/AVG button to cycle through measurement display modes:

  • Instantaneous Value: Displays the current real-time measurement.
  • Maximum Value (MAX): Displays the highest value recorded since power-on or reset.
  • Average Value (AVG): Displays the average value recorded over time.
  • Maximum Average Value (MAX AVG): Displays the maximum of the averaged values.

5.3. Aṣayan Unit

Tẹ awọn UNIT/ENTER button to change the measurement unit. Available units include:

  • mV/m (millivolts per meter)
  • V/m (volts per meter)
  • µA/m (microamperes per meter)
  • mA/m (milliamperes per meter)
  • µW/m² (microwatts per square meter)
  • mW/m² (milliwatts per square meter)
  • µW/cm² (microwatts per square centimeter)

5.4. Data Hold (HOLD/ALARM)

Tẹ awọn HOLD/ALARM bọ́tìnì náà fún ìgbà díẹ̀ láti dì kíkà tó wà lórí ìbòjú náà. Tẹ̀ lẹ́ẹ̀kan síi láti tú u sílẹ̀.

5.5. Iṣẹ-ṣiṣe itaniji

Lati ṣeto ẹnu-ọna itaniji:

  1. Tẹ mọlẹ HOLD/ALARM button for approximately 2 seconds to enter alarm setting mode.
  2. Lo awọn MAX/AVG ati REC/TIME buttons to adjust the threshold value.
  3. Tẹ UNIT/ENTER lati jẹrisi ati fi eto pamọ.
  4. The audible buzzer alarm will activate if the measured value exceeds the set threshold.

5.6. Data Memory (XYZ/MEM)

The device supports manual storage of up to 200 data sets.

  1. To save a reading, press the XYZ/MEM button briefly. The current reading will be stored.
  2. To recall stored data, press and hold the XYZ/MEM button for approximately 2 seconds. Use the MAX/AVG ati REC/TIME buttons to navigate through stored records.
  3. Tẹ UNIT/ENTER to exit data recall mode.

5.7. Calibration Factor (CAL) Adjustment

The calibration factor can be adjusted if necessary. Refer to the SET button and its associated menu for this advanced setting. This feature allows for fine-tuning the meter's response.

5.8. Agbara Aifọwọyi

The device has an adjustable auto power-off function. The default time is 15 minutes. To adjust this setting:

  1. Tẹ mọlẹ SET bọtini lati tẹ awọn eto akojọ.
  2. Navigate to the auto power-off option using the MAX/AVG or REC/TIME awọn bọtini.
  3. Lo awọn MAX/AVG ati REC/TIME buttons to set the desired time (0 to 99 minutes).
  4. Tẹ UNIT/ENTER to confirm and save. Setting to 0 disables auto power-off.

6. Itọju

  • Nu ẹrọ naa pẹlu asọ ti o tutu, ti o gbẹ. Ma ṣe lo abrasive ose tabi epo.
  • Tí a kò bá lo ẹ̀rọ náà fún ìgbà pípẹ́, yọ bátìrì náà kúrò kí ó má ​​baà jẹ́ kí ó jò.
  • Tọju ẹrọ naa ni itura, aye gbigbẹ, kuro lati orun taara ati awọn iwọn otutu to gaju.

7. Laasigbotitusita

  • Ẹrọ ko ni agbara lori: Ṣàyẹ̀wò bí batiri ṣe ń ṣiṣẹ́ kí o sì rí i dájú pé batiri náà ní agbára tó. Rọ́pò rẹ̀ tí ó bá pọndandan.
  • Àwọn ìwé kíkà tí kò báramu: Ensure the device is held steady and away from large metallic objects or other strong RF sources that might interfere with measurements. Verify the selected unit is appropriate for the measurement range.
  • Itaniji ko dun: Check if the alarm function is enabled and the threshold is set correctly.
  • Ifihan ti ṣofo tabi baibai: Rọpo batiri naa.

8. Awọn pato

Ẹya ara ẹrọSipesifikesonu
Iwọn Igbohunsafẹfẹ50MHz to 3.5GHz
IfihanLiquid-crystal (LCD), 4-1/2 digits, max reading 19999
Orisi wiwọnDigital, Triaxial, Isotropic
Ipinnu Ifihan0.1mV/m, 0.1µA/m, 0.001µW/m², 0.001µW/cm²
Awọn ẹyamV/m, V/m, µA/m, mA/m, µW/m², mW/m², µW/cm²
Itaniji NgbohunBuzzer, adjustable threshold
Eto AkokoTypically 1.5s (0 to 90% measurement value)
Sample Oṣuwọn1.5 igba fun keji
Memory data200 data sets (manual)
Calibration Factor (CAL)adijositabulu
Batiri Iru9V NEDA 1604, IEC 6F22 or JIS 006P
Igbesi aye batiriIsunmọ awọn wakati 15
Agbara Aifọwọyi PaaDefault 15 minutes, adjustable 0~99 minutes
Awọn iwọn otutu ti nṣiṣẹ0°C si +50°C (32°F si 122°F)
Ọriniinitutu ti nṣiṣẹ25% si 75% RH
Ibi ipamọ otutu-10°C si +60°C (14°F si 140°F)
Ọriniinitutu ipamọ0% si 80% RH
Awọn iwọn6.3 x 7.68 x 2.36 inches (160 x 195 x 60 mm)
Iwọn0.44 Pounds (approx. 7 ounces / 200g)

9. Atilẹyin ọja ati Support

For warranty information and technical support, please refer to the documentation provided with your purchase or contact TENMARS customer service. Keep your purchase receipt as proof of purchase.

Awọn iwe aṣẹ ti o jọmọ - TM-195

Ṣaajuview TENMARS TM-192 / TM-192D Afọwọṣe olumulo Mita aaye Triaxial ELF oofa
Itọsọna olumulo fun TENMARS TM-192 ati TM-192D Triaxial ELF Mita aaye oofa. Iwe yii ṣe alaye awọn ẹya ọja, awọn ohun elo, awọn ilana wiwọn, awọn pato, awọn iṣọra ailewu, ati fifi sori ẹrọ sọfitiwia.
Ṣaajuview TENMARS TM-195 RF Field Strength Meter User Manual
User manual for the TENMARS TM-195 RF three-Axis Field Strength Meter, covering specifications, operation, measurements, and safety information.
Ṣaajuview TENMARS Lux/Fc Afọwọṣe Olumulo Mita Mita - Awọn awoṣe TM-201, TM-202, TM-204, TM-205, TM-205N
Itọsọna olumulo fun TENMARS Lux/Fc Light Mita (TM-201, TM-202, TM-204, TM-205, TM-205N). Iwe yii n pese alaye alaye lori apejuwe ọja, awọn iṣọra ailewu, awọn ẹya, awọn pato, awọn ilana ṣiṣe, apejuwe irinse, ifamọ ojulumo ojulumo, awọn aaye akiyesi, awọn ipele itanna ti a ṣeduro, ati rirọpo batiri.
Ṣaajuview TENMARS TM-102/TM-103 Sound Level Meter Datalogging User Manual
Comprehensive user manual for the TENMARS TM-102 and TM-103 Sound Level Meters, covering specifications, operation, calibration, measurement, and software usage for accurate sound level monitoring.
Ṣaajuview Tenmars TM-720 LUX/FC Light Meter User Manual
User manual for the Tenmars TM-720 LUX/FC Light Meter, detailing its features, specifications, operation, safety precautions, recommended illumination levels, and maintenance.
Ṣaajuview Tenmars TM-197 AC/DC oofa aaye Mita olumulo Afowoyi
Itọsọna olumulo pipe fun Tenmars TM-197 AC / DC Magnetic Field Mita, ṣe alaye awọn ẹya rẹ, iṣẹ ṣiṣe, awọn pato, ati itọju. Bo wiwọn ti awọn aaye oofa, idaduro tente oke, gige data, ati awọn ilana iṣeto.