Greisinger GPHU 014 MP/BNC pH Measurement Converter Instruction Manual

Model: GPHU 014 MP/BNC

1. Ifihan

This manual provides comprehensive instructions for the safe and efficient operation of your Greisinger GPHU 014 MP/BNC pH Measurement Converter. Please read this manual thoroughly before using the device to ensure proper functionality and to prevent damage. Keep this manual for future reference.

The GPHU 014 MP/BNC is designed to convert pH electrode signals into a standardized output, allowing for precise pH monitoring in various applications. It is intended for use with compatible pH electrodes featuring a BNC connector.

2. Alaye Aabo

  • Always operate the device within its specified environmental conditions.
  • Ma ṣe fi ẹrọ naa han si awọn iwọn otutu to gaju, imọlẹ orun taara, tabi ọriniinitutu giga.
  • Yago fun sisọ silẹ tabi tẹriba ẹrọ si awọn ipa to lagbara.
  • Ensure proper electrical connections are made according to the instructions to prevent electrical shock or damage.
  • Ma ṣe gbiyanju lati ṣii tabi tun ẹrọ naa funrararẹ. Tọkasi gbogbo iṣẹ si oṣiṣẹ oṣiṣẹ.
  • Use only compatible pH electrodes and accessories with this converter.
  • Dispose of the device and any associated components in accordance with local environmental regulations.

3. Package Awọn akoonu

Jọ̀wọ́ ṣàyẹ̀wò ohun tó wà nínú àpótí náà nígbà tí o bá gbà á. Tí àwọn nǹkan kan bá sọnù tàbí tí wọ́n bá bàjẹ́, kàn sí olùpèsè rẹ lẹ́sẹ̀kẹsẹ̀.

  • Greisinger GPHU 014 MP/BNC pH Measurement Converter
  • Ilana Afowoyi (iwe yii)
  • Mounting accessories (if applicable, check packaging)

Note: A pH electrode is not included with this converter and must be purchased separately.

4. Ọja Ipariview

The Greisinger GPHU 014 MP/BNC is a compact and robust pH measurement converter designed for industrial and laboratory applications. It features a clear digital display and intuitive controls for easy operation.

Greisinger GPHU 014 MP/BNC pH Measurement Converter front view

Nọmba 1: Iwaju view of the Greisinger GPHU 014 MP/BNC pH Measurement Converter. The display shows "7.46 pH" and includes "ATC" and "Temp" indicators. Below the display are three buttons labeled "SET 1", "▼ 2", and "▲ 3". A BNC connector is visible on the right side for electrode connection. Two small ports are visible at the bottom.

Greisinger GPHU 014 MP/BNC pH Measurement Converter side view

Nọmba 2: Apa view of the Greisinger GPHU 014 MP/BNC pH Measurement Converter, highlighting the BNC connector and the robust housing. The connector is labeled "AMPHENOL".

Awọn eroja pataki:

  • Ifihan Digital: Shows pH value, temperature, and status indicators.
  • ṢE ÀTẸ̀ṢẸ̀ Bọ́tìnì (1): A lo lati tẹ ati jẹrisi awọn eto.
  • Down Button (2): Used to navigate menus and decrease values.
  • Up Button (3): Used to navigate menus and increase values.
  • Asopọmọra BNC: For connecting a compatible pH electrode.
  • Iṣagbewọle agbara: (Location not visible in images, refer to device for exact position)

5. Eto

5.1. Iṣagbesori

The converter can be mounted using the integrated mounting points (screws visible in Figure 1). Ensure the mounting location is stable, free from excessive vibration, and within the specified operating temperature range.

5.2. Connecting the pH Electrode

  1. Ensure the converter is powered off.
  2. Carefully connect your pH electrode's BNC connector to the BNC input on the right side of the GPHU 014 MP/BNC converter (refer to Figure 1 and 2).
  3. Ensure the connection is secure but do not overtighten.

5.3. Asopọ agbara

Connect the appropriate power supply to the converter. Refer to the device labeling for specific voltage and current requirements. Ensure the power source is stable and matches the device's specifications.

5.4. Iṣatunṣe ibẹrẹ

Before first use, or after prolonged storage, the pH converter and electrode system must be calibrated. Refer to your pH electrode's manual for specific calibration buffer requirements.

  1. Agbara lori ẹrọ naa.
  2. Access the calibration menu using the SET button. (Specific steps may vary, consult the device's on-screen prompts or a more detailed technical manual if available).
  3. Follow the on-screen instructions to perform a 2-point or 3-point calibration using standard pH buffer solutions (e.g., pH 4.01, 7.00, 10.01).
  4. Rinse the electrode thoroughly with distilled water between buffer solutions.
  5. Confirm calibration settings.

6. Ṣiṣẹ

6.1. Gbigbe Iwọn kan

  1. Ensure the pH electrode is properly connected and calibrated.
  2. Immerse the pH electrode into the sample solution, ensuring the sensing bulb and reference junction are fully submerged.
  3. Allow sufficient time for the reading to stabilize on the digital display. The display will show the current pH value.
  4. Note any temperature readings if the electrode has an integrated temperature sensor and ATC (Automatic Temperature Compensation) is active.

6.2. Lilọ kiri Akojọ aṣyn

The SET, ▼, and ▲ buttons are used to navigate through the device's menu options. Press SET to enter the menu or confirm a selection. Use ▼ and ▲ to scroll through options or adjust values.

  • SET (1): Enter/Exit menu, confirm selection.
  • ▼ (2): Scroll down, decrease value.
  • ▲ (3): Scroll up, increase value.

Specific menu functions (e.g., setting alarm limits, output configuration) will be detailed in a more advanced technical manual or on-screen prompts.

7. Itọju

7.1. Cleaning the Converter

Wipe the exterior of the converter with a soft, damp cloth. Do not use abrasive cleaners or solvents. Ensure no liquid enters the device.

7.2. Itoju pH Electrode

Proper maintenance of the pH electrode is crucial for accurate readings and extended lifespan. Refer to your specific pH electrode's instruction manual for detailed care instructions. General guidelines include:

  • Always keep the electrode's sensing bulb hydrated, typically in a storage solution or pH 7 buffer. Never store dry.
  • Rinse the electrode with distilled or deionized water after each measurement.
  • Clean the electrode regularly to remove sample residues. Specific cleaning solutions may be required depending on the application.
  • Recalibrate the electrode regularly, especially if readings become unstable or inaccurate.

7.3. Ibi ipamọ

When not in use for extended periods, store the converter in a clean, dry environment at room temperature. Ensure the pH electrode is properly stored according to its manufacturer's recommendations.

8. Laasigbotitusita

Isoro Owun to le Fa Ojutu
No display/Device not powering on No power, faulty power supply, incorrect connection. Check power cable and source. Ensure correct voltage.
Unstable or inaccurate pH readings Electrode not calibrated, dirty electrode, damaged electrode, air bubbles, incorrect temperature compensation. Recalibrate electrode. Clean electrode. Replace electrode if damaged. Ensure electrode is fully submerged. Check ATC settings.
"Error" message on display Electrode fault, out of range measurement, internal device error. Check electrode connection. Try another electrode if available. Refer to specific error codes in a detailed technical manual. Contact support if persistent.
O lọra esi akoko Aged or dirty electrode, low sample iwọn otutu. Clean or recondition electrode. Allow electrode to equilibrate with sample temperature. Consider replacing electrode.

If the problem persists after attempting these solutions, please contact Greisinger customer support or your local distributor.

9. Awọn pato

Ẹya ara ẹrọ Ẹ̀kúnrẹ́rẹ́
Olupese GREISINGER
Nọmba Awoṣe Nkan 602331
Awọn iwọn Ọja (L x W x H) 5.5 x 8.2 x 8 cm; 9.07 giramu
ASIN B00ICKLET6
Opoiye ti awọn ohun kan 1
Battery(ies) included Rara
Battery(ies) required Rara
Ọjọ́ tí a kọ́kọ́ rí lórí Amazon.com.be Oṣu Kẹjọ Ọjọ 12, Ọdun 2022
Olùpèsè dáwọ́ dúró Rara
Apoju awọn ẹya wiwa Alaye ko si lori apoju awọn ẹya ara

10. Atilẹyin ọja ati Support

Greisinger products are manufactured to high-quality standards. For information regarding warranty terms and conditions, please refer to the warranty card included with your product or visit the official Greisinger webojula.

For technical support, troubleshooting assistance beyond this manual, or to inquire about spare parts and service, please contact your authorized Greisinger dealer or the manufacturer directly. Contact information can typically be found on the product packaging or the official Greisinger webojula.

Awọn iwe aṣẹ ti o jọmọ - GPHU 014 MP/BNC

Ṣaajuview GREISINGER GE xxx Ilana Ṣiṣẹ: pH Electrodes
Iwe afọwọkọ iṣiṣẹ okeerẹ fun awọn amọna pH jara GREISINGER GE xxx, aabo ibora, awọn ipilẹ, wiwọn, isọdiwọn, itọju, yiyan, ati awọn pato. Pẹlu itọnisọna alaye fun wiwọn pH deede.
Ṣaajuview GREISINGER EBUW 232 EASYBUS Protocol Converter Operating Manual
Operating manual for the GREISINGER EBUW 232 EASYBUS protocol converter, detailing its specifications, general information, and safety requirements.
Ṣaajuview GREISINGER GIA 20 EB / PK Afowoyi: Asopọmọra ati isẹ
Iwe afọwọkọ yii n pese awọn itọnisọna okeerẹ fun ifihan iṣakoso microprocessor GREISINGER GIA 20 EB/PK, ibojuwo, ati ẹrọ iṣakoso. O ni wiwa awọn ilana aabo, awọn asopọ itanna fun ọpọlọpọ awọn ifihan agbara titẹ sii ati awọn iru iṣẹjade, iṣeto ẹrọ nipa lilo sọfitiwia, ibi-iyipada ati awọn eto itaniji, aiṣedeede / atunṣe ite, wiwo ni tẹlentẹle, awọn koodu aṣiṣe, awọn pato, ati awọn akọsilẹ isọnu.
Ṣaajuview GREISINGER GMH 3710 Precision Thermometer Operating Manual
This operating manual provides detailed instructions for the GREISINGER GMH 3710 precision thermometer. It covers intended use, safety precautions, product specifications, handling, configuration, output options, input adjustments, probe connections, basic temperature measurement principles, fault messages, and disposal guidelines for Pt100 temperature sensors.
Ṣaajuview Greisinger GDH 200 Series Manometers and Barometers: Technical Overview
Discover the Greisinger GDH 200 series handheld pressure measurement instruments. This document details the GDH 200-07 fine manometer, GDH 200-13 manometer, and GDH 200-14 vacuum/barometer, highlighting their specifications and features for professional use.
Ṣaajuview GREISINGER GMH 38-LW1 / -LW2 Moisture Meter Operating Manual Appendix
Operating manual appendix for the GREISINGER GMH 38-LW1 / -LW2 moisture measuring set for agricultural applications. Covers intended use, safety, product description, handling, principles of measurement, and specifications for wood chips, grain, hay, and straw.