Ọrọ Iṣaaju
The WTW Sentix H pH Electrode is specifically designed for challenging measurements, particularly in high alkaline solutions. This specialized electrode ensures accurate pH readings even in low-ion samples or when dealing with very small volumes. Its robust design and advanced features make it a reliable tool for various laboratory and industrial applications.

Nọmba 1: WTW Sentix H pH Electrode. This image displays the full length of the electrode, highlighting its slender body, the S7 plug head at the top, and the sensing bulb at the bottom. The blue labeling on the electrode body is also visible.
Ṣeto
Proper setup is crucial for optimal performance of your Sentix H pH Electrode. Follow these steps:
- Ṣii silẹ: Carefully remove the electrode from its packaging. Inspect for any visible damage.
- Storage Solution Removal: The electrode tip is typically protected by a cap containing a storage solution. Remove this cap. Do not discard the cap or solution, as it is needed for storage.
- Rinsin: Rinse the electrode tip thoroughly with distilled or deionized water to remove any residual storage solution. Gently blot dry with a lint-free tissue.
- Connecting to Meter: Connect the S7 plug head of the electrode to the corresponding S7 input port on your compatible pH meter. Ensure a secure connection.
- Pre-conditioning (if necessary): For new electrodes or those that have been stored dry, immerse the electrode tip in a suitable electrode storage solution or pH 7 buffer for at least 30 minutes to an hour before first use.
Awọn ilana Iṣiṣẹ
The Sentix H electrode is optimized for high alkaline solutions. Adhere to the following operational guidelines:
- Iṣatunṣe: Always calibrate the electrode with fresh pH buffer solutions (e.g., pH 7.00 and pH 10.00 or pH 12.00 for high alkaline range) before each measurement session or as recommended by your pH meter's manual. Ensure the buffers are at the same temperature as your samples.
- Sample Immersion: Immerse the electrode tip fully into the sample solution, ensuring the liquid junction is completely submerged. Stir the sample gently to ensure homogeneity and stable readings.
- Reading Stability: Allow sufficient time for the reading to stabilize. This electrode features special glass membranes and an adapted diaphragm technique for optimal accuracy, but stabilization time can vary depending on the sample.
- Rinsing Between Sample: Rinse the electrode thoroughly with distilled or deionized water between each sample measurement to prevent cross-contamination.
- Biinu iwọn otutu: Utilize the temperature compensation feature of your pH meter, or ensure samples and buffers are at a consistent temperature.
Itoju
Regular maintenance extends the lifespan and ensures the accuracy of your electrode:
- Ninu:
- Isọsọ gbogboogbo: After each use, rinse the electrode thoroughly with distilled or deionized water.
- Awọn idogo Protein: For protein contamination, soak the electrode in a pepsin/HCl solution for several hours.
- Inorganic Deposits: For inorganic deposits, soak in a 0.1 M HCl or 0.1 M HNO3 solution for 15-30 minutes.
- Grease/Oil: For grease or oil, clean with a mild detergent solution, then rinse thoroughly.
- Ibi ipamọ: Always store the electrode with its tip immersed in a suitable electrode storage solution (e.g., 3M KCl solution) or the original storage solution provided. Never store the electrode dry or in distilled water.
- Electrolyte Level: Ensure the electrolyte level in the electrode is maintained. Refill if necessary according to the electrode's specific requirements.
- Ayewo: Periodically inspect the glass bulb and junction for cracks, scratches, or blockages.
Laasigbotitusita
Tí o bá ní ìṣòro, wo tábìlì ní ìsàlẹ̀ yìí fún àwọn ìṣòro àti ìdáhùn tó wọ́pọ̀:
| Isoro | Owun to le Fa | Ojutu |
|---|---|---|
| Unstable or Drifting Readings |
|
|
| Àkókò Ìdáhùn Díẹ̀díẹ̀ |
|
|
| Inaccurate Readings / Calibration Failure |
|
|
Awọn pato
| Iwa | Ẹ̀kúnrẹ́rẹ́ |
|---|---|
| Orukọ awoṣe | Sentix H pH Electrode |
| Nọmba Awoṣe Nkan | 0916A26EA |
| Olupese | WTW |
| ASIN | B01N7MVTLV |
| Ohun elo | High Alkaline Solutions, Low-ion Samples, Small Volumes |
| Pulọọgi Ori | S7 |
| Key Awọn ẹya ara ẹrọ | Special glass membranes, adapted diaphragm technique |
Atilẹyin ọja ati Support
For information regarding warranty coverage, technical support, or service, please contact WTW directly or refer to their official website. Keep your purchase receipt and model number handy for faster assistance.
Olupese: WTW