Apera Instruments AI532

Apera Instruments EC820 Benchtop Conductivity Meter Kit

Itọsọna olumulo

1. Ifihan ati Ọja Loriview

The Apera Instruments EC820 Benchtop Conductivity Meter is a high-precision instrument designed for accurate measurement of Conductivity (EC), Total Dissolved Solids (TDS), Salinity, Resistivity, and temperature. This meter is equipped with advanced features for reliable and stable readings across a wide range of applications.

This manual provides detailed instructions for the proper setup, operation, maintenance, and troubleshooting of your EC820 meter to ensure optimal performance and longevity.

Apera Instruments EC820 Benchtop Conductivity Meter Kit with electrode and beaker

Figure 1: Apera Instruments EC820 Benchtop Conductivity Meter Kit. The image shows the main meter unit, a flexible electrode arm and holder, and the conductivity electrode submerged in a beaker of liquid.

2. Awọn iṣọra aabo

  • Always handle the meter and electrodes with care to prevent damage.
  • Do not immerse the main meter unit in water. The unit is IP54 water-resistant, but not submersible.
  • Lo oluyipada agbara ti a pese nikan.
  • Keep calibration solutions and samples away from children and pets. Follow safety data sheets (SDS) for handling specific chemicals.
  • Ensure the meter is placed on a stable, level surface during operation.
  • Ge asopọ agbara ṣaaju ṣiṣe mimọ tabi ṣiṣe itọju.

3. Package Awọn akoonu

Daju pe gbogbo awọn nkan ti a ṣe akojọ rẹ si isalẹ wa ninu package rẹ:

  • EC820 konge Benchtop Conductivity Mita
  • 2401T-F Platinum Glass-body Conductivity Electrode
  • Flexible Electrode Arm and Holder
  • Okun USB
  • PC-Link Software Flash Drive
  • Power Adaptor (Input: 120V AC 60Hz; Output: 9V 300mA)
  • Calibration Solutions:
    • 84 µS/cm (1 vial)
    • 1413 µS/cm (1 vial)
    • 12.88 mS/cm (1 vial)
Apera Instruments 2401T-F Platinum Glass-body Conductivity Electrode with cables

Figure 2: The 2401T-F Platinum Glass-body Conductivity Electrode, showing its connection cables.

4. Eto

  1. Unpack the Meter: Ni ifarabalẹ yọ gbogbo awọn paati kuro ninu apoti.
  2. Assemble Electrode Arm: Attach the flexible electrode arm to the meter unit. Ensure it is securely fastened.
  3. Connect Electrode: Plug the 2401T-F Conductivity Electrode into the designated "COND" port on the back of the meter. Ensure the connection is firm.
    Pada view of Apera Instruments EC820 meter showing connection ports

    Figure 3: Rear panel of the EC820 meter, highlighting the "COND" and "TEMP" ports, along with RS232 and DC 9V power input.

  4. So agbara: Plug the power adapter into the "DC 9V" port on the back of the meter, then connect it to a standard 120V AC 60Hz power outlet.
  5. Agbara Ibẹrẹ Tan: Press the power button to turn on the meter. The LCD display should illuminate.

5. Awọn ilana Iṣiṣẹ

5.1. Iṣatunṣe

The EC820 meter features 1 to 4 points auto-calibration. Regular calibration ensures accurate measurements.

  1. Prepare Solutions: Use the provided 84 µS/cm, 1413 µS/cm, and 12.88 mS/cm calibration solutions. Ensure solutions are at room temperature.
  2. Fi omi ṣan elekitirodu: Rinse the conductivity electrode thoroughly with distilled or deionized water before and after each measurement or calibration. Gently blot dry with a lint-free tissue.
  3. Tẹ Ipo Iṣatunṣe: With the meter on, press the CAL button. The display will show "CAL" and prompt for the first calibration point.
  4. First Point Calibration: Immerse the electrode into the 1413 µS/cm standard solution. Wait for the reading to stabilize (indicated by a stable icon on the display). The meter will automatically recognize the solution. Press WOLE lati jẹrisi.
  5. Subsequent Points (Optional): For higher accuracy, proceed with 84 µS/cm and 12.88 mS/cm solutions following the same procedure. The meter will guide you through the process.
  6. Exit Calibration: After completing desired calibration points, press the MEAS button to return to measurement mode.

Akiyesi: The meter has a self-diagnosis function to help identify and fix common calibration problems. Refer to the on-screen prompts.

5.2. Wiwọn

  1. Fi omi ṣan elekitirodu: Fi omi ṣan elekiturodu pẹlu omi distilled.
  2. Ẹ̀rọ ìtẹ̀mọ́lẹ̀: Fi elekitirodu sinu sample solution, ensuring the sensor is fully submerged.
  3. Wait for Stability: Allow the reading to stabilize. The stable reading icon will appear on the LCD.
  4. Record Reading: Read the Conductivity (EC), TDS, Salinity, or Resistivity value displayed on the large LCD. Temperature will also be displayed simultaneously.
  5. Yi Awọn Ẹka pada: Tẹ awọn UNIT button to cycle through different measurement units (e.g., µS/cm, mS/cm, ppm, ppt).
  6. Ibi ipamọ data: The meter supports GLP data management with 500 groups of data storage. Use the M+ / R-M buttons to store and recall data.
  7. Gbejade Data: Connect the meter to a PC via the USB cable and use the provided PC-Link software (Windows-based system only) to export data.

6. Itọju

  • Ninu Electrode:
    • Regularly rinse the electrode with distilled or deionized water.
    • For stubborn deposits, refer to the electrode's specific cleaning instructions. Generally, a mild detergent solution or a specific electrode cleaning solution can be used, followed by thorough rinsing.
    • Do not scratch or damage the platinum black sensor.
  • Itọju itanna:
    • Always store the conductivity electrode in its protective cap filled with a proper electrode storage solution (e.g., 3M KCl solution or Apera Instruments storage solution). Never store dry.
  • Ìmọ́tótó Mita: Wipe the main meter unit with a clean, damp asọ. Ma ṣe lo abrasive ose tabi epo.
  • Awọn imudojuiwọn sọfitiwia: Check the Apera Instruments website periodically for any software updates for the PC-Link software.
Close-up of the Apera Instruments conductivity electrode tip

olusin 4: Isunmọ view of the platinum glass-body conductivity electrode tip, which requires careful handling and proper storage.

7. Laasigbotitusita

IsoroOwun to le FaOjutu
Mita naa ko ṣiṣẹKò sí ìpèsè agbára; adapter tí kò ní àbùkù; ìsopọ̀ tí kò ní àbùkù.Check power adapter connection and outlet. Ensure adapter is functional.
Awọn kika aiduroDirty electrode; air bubbles on sensor; incorrect calibration; temperature fluctuations.Clean electrode; gently tap electrode to remove bubbles; recalibrate meter; ensure sample temperature is stable.
Aṣiṣe odiwọnContaminated calibration solution; incorrect solution used; damaged electrode.Use fresh calibration solutions; ensure correct solution for calibration point; inspect electrode for damage.
No data transfer to PCIncorrect USB connection; software not installed or running; driver issues.Verify USB cable connection; ensure PC-Link software is installed and open; check device manager for driver issues.

8. Awọn pato

ParamitaIye
Nọmba awoṣeAI532
Idiwọn ParametersConductivity (EC), TDS, Salinity, Resistivity, Temperature
Ibiti conductivity0 to 2,000 mS
Yiye± 0.5% FS
Idiwọn Points1 to 4 points auto-calibration (84µS/cm, 1413µS/cm, 12.88mS/cm, 111.8mS/cm)
Biinu iwọn otutuBiinu Igba otutu Laifọwọyi (ATC)
Ibi ipamọ data500 groups (GLP compliant)
AsopọmọraUSB for PC data export
Ifihan IruTobi Llit Llit
Ibi ti ina elekitiriki ti nwa9V DC, 300mA (120V AC 60Hz adapter included)
Idaabobo IngressIP54 Omi-sooro
Awọn iwe-ẹriCE, RoHS, ISO 9001:2015
Package Mefa10.63 x 9.88 x 6.61 inches
Iwọn6.5 iwon

9. Atilẹyin ọja ati Support

Apera Instruments products are manufactured under an ISO 9001:2015 certified quality system, ensuring high standards of quality and reliability. For specific warranty information and technical support, please refer to the warranty card included with your product or visit the official Apera Instruments webojula.

For assistance with troubleshooting, calibration, or general inquiries, please contact Apera Instruments customer service. Contact details can typically be found on their official webaaye tabi ni apoti ọja.

Awọn orisun Ayelujara: Apera Instruments Store on Amazon

Awọn iwe aṣẹ ti o jọmọ - AI532

Ṣaajuview Apera EC20 Apo Idanwo Conductivity: Ilana itọnisọna & Itọsọna
Iwe itọnisọna pipe fun Apera EC20 Oluṣeto Imudaniloju Apo. Kọ ẹkọ nipa iṣeto, isọdọtun, wiwọn adaṣe, awọn pato imọ-ẹrọ, ati alaye atilẹyin ọja.
Ṣaajuview APERA INSTRUMENTS PC950 Benchtop pH/Conductivity Meter Instruction Manual
Comprehensive instruction manual for the APERA INSTRUMENTS PC950 Benchtop pH/Conductivity Meter, covering specifications, operation, calibration, maintenance, and troubleshooting for laboratory use.
Ṣaajuview Apera Instruments EC400S Portable Conductivity Meter Instruction Manual
Comprehensive instruction manual for the Apera Instruments EC400S Portable Conductivity Meter, covering features, specifications, operation, calibration, sample measurement, maintenance, parameter settings, and USB data communication.
Ṣaajuview Apera Instruments SX731 Portable pH/Conductivity/ORP Meter User Manual
User manual for the Apera Instruments SX731 Portable pH/Conductivity/ORP Meter, covering its features, technical specifications, operation, calibration, maintenance, and warranty.
Ṣaajuview Apera SX823-B pH/Conductivity Portable Meter Instruction Manual
This instruction manual provides detailed information on the Apera SX823-B portable pH/Conductivity meter, covering its features, specifications, operation, calibration procedures, maintenance, and troubleshooting.
Ṣaajuview Apera Instruments EC20 Pocket Conductivity Tester Ilana itọnisọna
Iwe afọwọkọ olumulo fun Apera Instruments EC20 Oluyẹwo Iwa Apo, fifi sori batiri ni wiwa, awọn iṣẹ bọtini foonu, isọdiwọn, wiwọn adaṣe, awọn pato imọ-ẹrọ, ati alaye atilẹyin ọja.