Thank you for choosing the PerfectPrime TC0521 4-Channel Thermocouple Thermometer. This instrument is designed for precise temperature measurement across multiple points simultaneously, supporting K, J, E, T, N, R, and S type thermocouples. It features data logging capabilities, USB PC connectivity, and wireless mobile app integration for convenient monitoring and analysis. This manual provides essential information for the safe and efficient operation of your device.
2. Ọja Ipariview
Key Awọn ẹya ara ẹrọ
- Four Channels Thermocouple Thermometer supporting K, J, E, T, N, R, S Type.
- 3.35" LCD display for clear readings.
- Internal DataLog Storage: 32,000 data logger records per channel.
- USB PC Interface: Real-time monitoring, graphic display, data logging, storage, and data import/export.
- Wireless connectivity to mobile apps (iOS and Android).
- Fast response and sampoṣuwọn ling.
- Adjustable auto power-off timer.
- Temperature alarm function.
- Instant recall function for stored data.
Idanimọ paati



3. Eto
3.1 Fifi sori batiri
- Wa yara batiri ni ẹhin ẹrọ naa.
- Ṣii ideri batiri naa.
- Insert 4 AAA 1.5V batteries, ensuring correct polarity.
- Pa ideri batiri naa ni aabo.
3.2 Àwọn Ìwádìí Thermocouple Tó Sopọ̀
- Identify the four input ports (T1, T2, T3, T4) at the top of the thermometer.
- Connect the K-type bead probes (or other compatible thermocouple types) to the desired input ports. Ensure the positive (+) and negative (-) terminals of the probe connector align with the corresponding markings on the thermometer.
- For accurate readings, ensure the probes are securely inserted.

3.3 Software and Mobile App Installation
- Sọfitiwia PC: Download the Windows software (compatible with XP/Vista/7/8/10) from the manufacturer's webojula. Tẹle awọn ilana loju iboju fun fifi sori.
- Ohun elo Alagbeka: Wa fun the PerfectPrime mobile app on the iOS App Store or Android Play Store. Install the app and follow the in-app instructions to connect wirelessly via Bluetooth.
4. Awọn ilana Iṣiṣẹ
4.1 Agbara Tan / Pa
- Tẹ awọn Agbara bọtini lati tan thermometer naa.
- Tẹ mọlẹ Agbara button again to turn the thermometer off.
- The device features an adjustable auto power-off timer to conserve battery life.
4.2 Ìwọ̀n Ìwọ̀n Òtútù Pàtàkì
- Ensure probes are correctly connected to the input ports.
- Power on the device. The display will show temperature readings for each active channel.
- Place the probe tips at the desired measurement points. Allow a few moments for the readings to stabilize.
4.3 Unit Selection (°C/°F)
- Tẹ awọn UNIT button to toggle between Celsius (°C) and Fahrenheit (°F) temperature units.
4.4 Data idaduro Išė
- Tẹ awọn DIMU button to freeze the current temperature readings on the display.
- Tẹ awọn DIMU button again to release the hold and resume live readings.
4.5 Min/Max/Avg Functions
- Tẹ awọn MAX button to display the maximum temperature recorded since the device was powered on or the function was reset.
- Tẹ awọn MIN button to display the minimum temperature recorded.
- Tẹ awọn AVG button to display the average temperature recorded.
- To reset these values, typically a long press of the respective button or a specific sequence might be required (refer to the included mini-manual for exact steps).
4.6 Imọlẹ ẹhin
- Tẹ awọn Imọlẹ ẹhin button (often represented by a light bulb icon) to turn the display backlight on or off for improved visibility in low-light conditions.
4.7 Data Logging and Wireless Connectivity
- The device can store up to 32,000 data records per channel internally.
- Connect the thermometer to a PC via the micro USB cable or wirelessly to your mobile device via Bluetooth.
- Use the dedicated PC software or mobile app to configure data logging intervals, view real-time data, analyze trends, and export recorded data.
5. Itọju
- Ninu: Mu ẹrọ naa nu pẹlu asọ ti o gbẹ. Ma ṣe lo abrasive ose tabi epo.
- Rirọpo Batiri: Rọpo awọn batiri nigbati ifihan batiri kekere ba han lori ifihan lati rii daju pe o n ṣiṣẹ nigbagbogbo.
- Ibi ipamọ: Store the thermometer in a cool, dry place away from direct sunlight and extreme temperatures when not in use. Remove batteries if storing for extended periods.
- Itọju Iwadii: Handle thermocouple probes carefully. Avoid bending or kinking the wires excessively. Replace damaged probes to maintain accuracy.
6. Laasigbotitusita
- Ko si Agbara: Ṣayẹwo fifi sori batiri ati rii daju pe awọn batiri jẹ alabapade.
- Awọn kika ti ko pe: Verify probe connections, ensure the correct thermocouple type is selected (if applicable), and check for damaged probes. Allow sufficient time for readings to stabilize.
- Awọn oran Asopọmọra: For PC connection, ensure the USB cable is properly connected and drivers are installed. For mobile app, ensure Bluetooth is enabled on both devices and the app is correctly paired.
- Awọn aṣiṣe ifihan: If the display shows '----' or other error messages, check probe connections and ensure the measured temperature is within the device's specified range.
7. Awọn pato
| Ẹya ara ẹrọ | Sipesifikesonu |
|---|---|
| Measurement Temperature Range (K-type) | -200°C si 1372°C (-328°F si 2501°F) |
| Measurement Temperature Range (J-type) | -200°C si 1000°C (-328°F si 1832°F) |
| Measurement Temperature Range (E-type) | -200°C si 750°C (-328°F si 1382°F) |
| Measurement Temperature Range (T-type) | -200°C si 400°C (-328°F si 752°F) |
| Accuracy (J, K, E, T, N-type) | ±(0.1% of reading + 0.7°C) above -100°C; ±(0.5% of reading + 0.7°C) below -100°C |
| Accuracy (R, S-type) | ±(0.2% ti kika + 1.4°C) |
| Temperature Resolution (K, J, E, T, N-type) | 0.1°C (<600°C), 1°C (≥600°C) |
| Temperature Resolution (R, S-type) | 0.2°C (<600°C), 1°C (≥600°C) |
| Sample Oṣuwọn | 2 igba fun keji |
| Batiri Iru | 4 x AAA 1.5V awọn batiri |
| Battery Lifetime (Bluetooth OFF) | Isunmọ. Awọn wakati 120 (batiri ipilẹ) |
| Battery Lifetime (Bluetooth ON) | Isunmọ. Awọn wakati 30 (batiri ipilẹ) |
| Isẹ otutu | 0°C si 50°C (32°F si 122°F) |
| Ọriniinitutu isẹ | 10 si 90% RH (ti kii ṣe itọlẹ) |
| Ibi ipamọ otutu | -20°C si 60°C (-4°F si 140°F) |
| Ọriniinitutu ipamọ | 10 si 75% RH |
| Awọn iwọn | 187mm (L) × 75mm (W) × 29mm (H) |
| Iwọn | Isunmọ. 290g |
| Ifihan Iru | LCD |
| Asopọmọra Technology | USB, Bluetooth |
| Ohun elo ita | Polycarbonate (PC) |
8. Ohun ti o wa ninu Apoti
- PerfectPrime TC0521 Thermocouple Thermometer
- Ilana Afowoyi (iwe yii)
- AAA 1.5V batteries (4 pcs)
- Okun USB Micro
- K-type bead probes (2 pcs)
- Windows software (available for download)
9. Atilẹyin ọja ati Support
While specific standard warranty details are not provided in the product information, PerfectPrime is committed to product quality. An optional 1-Year Extended Warranty Plan by Onsitego is available for purchase. For technical support, troubleshooting assistance, or warranty inquiries, please refer to the contact information provided with your purchase or visit the official PerfectPrime webojula.
This device has a Certificate of Calibration from the National Measurement Laboratory R.O.C (NML/ROC) and is Made in Taiwan, ensuring quality and reliability.