1. Ifihan
The Fluke 1507 Digital Megohmmeter Insulation Resistance Tester is a robust and reliable instrument designed for comprehensive insulation testing in industrial and electrical environments. This manual provides essential information for the safe and effective operation, maintenance, and troubleshooting of your Fluke 1507 tester.
Key features include multiple insulation test voltages, automatic calculation of Polarization Index (PI) and Dielectric Absorption Ratio (DAR), a Compare (Pass/Fail) function for repetitive testing, and live circuit detection for enhanced safety.
2. Alaye Aabo
WARNING: Always read and understand all safety information before operating the Fluke 1507 tester. Failure to do so may result in serious injury or death.
- Wiwa Circuit Live: The tester is equipped with live circuit detection, which prevents insulation tests if a voltage greater than 30 Volts is detected. This feature is crucial for user protection.
- Auto-Discharge: The 1507 provides auto-discharge of capacitive voltage for added user protection after testing.
- Ẹka Iwọn: This instrument complies with ANSI/ISA 82.02.01 (61010-1) 2004, CAN/CSA-C22.2 No. 61010-1-04, and IEC/EN 61010-1 2nd edition for Measurement Category IV 600 V (CAT IV). Always use the tester within its specified measurement category.
- Lilo daradara: Use only the test leads and accessories supplied or approved by Fluke. Inspect test leads for damage before each use.
- Awọn ipo Ayika: Do not use the tester in wet or explosive environments. Ensure the operating area is dry and well-ventilated.
3. Components and Included Accessories
The Fluke 1507 Insulation Resistance Tester package typically includes the following items:
- Fluke 1507 Insulation Resistance Tester Unit
- Àwọn Ìdánwò TL224
- TP74 Test Probes
- Awọn agekuru Alligator
- Holster
- Iwadi Latọna jijin
- Four AA Batteries (pre-installed or included separately)

4. Eto
4.1. Fifi sori batiri
The Fluke 1507 requires four AA alkaline batteries for operation. These are typically included with the unit. To install or replace batteries:
- Ensure the tester is turned OFF.
- Wa yara batiri ni ẹhin ẹyọ naa.
- Lo screwdriver lati ṣii ideri iyẹwu batiri naa.
- Insert the four AA batteries, observing the correct polarity as indicated inside the compartment.
- Rọpo ideri kompaktimenti batiri ki o ni aabo pẹlu dabaru.
4.2. Nsopọ Awọn itọsọna Igbeyewo
Connect the test leads to the appropriate input jacks on the tester. For insulation resistance measurements, use the 'V Ω' and 'COM' jacks. Ensure connections are secure before proceeding with any tests.
5. Awọn ilana Iṣiṣẹ
5.1. Agbara lori / Paa
To power on the Fluke 1507, rotate the rotary switch from the 'OFF' position to any desired measurement function. The large, backlit display will illuminate. To power off, rotate the switch back to 'OFF'. The unit also features an auto-power off function to conserve battery life.

5.2. Yiyan Igbeyewo Voltages
The Fluke 1507 offers multiple insulation test voltages. Rotate the main rotary switch to the 'INSULATION' section. Then, use the dedicated voltage selection buttons (50 V, 100 V, 250 V, 500 V, 1000 V) to choose the appropriate test voltage fun ohun elo rẹ.
5.3. Performing Insulation Tests
Once the desired test voltage is selected and test leads are connected to the circuit under test:
- Tẹ mọlẹ Idanwo button to initiate the insulation test. The display will show the measured insulation resistance.
- For hands-free operation during repetitive tests, utilize the remote test probe.
- The live circuit detection feature will prevent a test if voltage exceeds 30V, indicated by a warning on the display.


5.4. Compare (Pass/Fail) Function
The Compare (Pass/Fail) function simplifies repetitive testing by allowing you to set a pass/fail threshold. Refer to the on-screen prompts and the user interface for setting this function. The tester will then indicate whether the measured resistance meets the predefined criteria.
5.5. Atọka Polarization (PI) ati Dielectric Absorption Ratio (DAR)
The Fluke 1507 automatically calculates Polarization Index (PI) and Dielectric Absorption Ratio (DAR), saving time and ensuring accurate assessment of insulation condition. These functions are typically accessed via dedicated buttons on the front panel.

6. Itọju
6.1. Ninu
To clean the exterior of the tester, use a damp cloth and a mild detergent. Do not use abrasive cleaners or solvents. Ensure the unit is completely dry before storage or next use.
6.2. Batiri Rirọpo
Replace batteries when the low battery indicator appears on the display. Refer to Section 4.1 for detailed battery installation instructions. Always use fresh alkaline batteries.
6.3. Iṣatunṣe
For optimal performance and accuracy, periodic calibration by an authorized service center is recommended. Refer to your warranty information or contact Fluke support for calibration schedules and services.
7. Laasigbotitusita
This section provides solutions to common issues you might encounter with your Fluke 1507 tester.
| Isoro | Owun to le Fa | Ojutu |
|---|---|---|
| Oluyẹwo ko ni agbara lori. | Awọn batiri ti o ku tabi ti ko tọ ti fi sori ẹrọ. | Ṣayẹwo polarity batiri ki o rọpo pẹlu awọn batiri AA tuntun. |
| Insulation test does not start. | Live voltage detected on the circuit. | Ensure the circuit is de-energized or voltage is below 30V. The tester will not perform an insulation test if live voltage is present for safety. |
| Awọn kika ti ko pe. | Damaged test leads or improper connection. | Inspect test leads for damage. Ensure leads are securely connected to the tester and the circuit. |
| Ifihan ti wa ni baibai tabi flicker. | Batiri kekere. | Rọpo awọn batiri. |
8. Awọn pato
| Sipesifikesonu | Iye |
|---|---|
| Brand | Fluke |
| Nọmba awoṣe | 1507 |
| Orisun agbara | Agbara Batiri (batiri AA 4) |
| Idanwo idabobo Voltages | 50 V, 100 V, 250 V, 500 V, 1000 V |
| Min. Ṣiṣẹ Voltage | 50 Volts (DC) |
| Voltage Rating | 600 Volts |
| Yiye wiwọn | +/- 0.5% |
| Iwọn Nkan | 2.2 poun (kilogram 1) |
| Awọn iwọn Ọja (L x W x H) | 11.77 x 15.2 x 3.07 inches |
| Ohun elo | Ṣiṣu |
| Àwọ̀ | Yellow |
| Ibamu | EN 61010-1, CAT IV 600 V |
| Awọn irinše to wa | Remote probe, test leads, test probes, alligator clips |
9. Atilẹyin ọja ati Support
9.1. atilẹyin ọja Information
The Fluke 1507 Digital Megohmmeter Insulation Resistance Tester comes with a 1-Year warranty. Please retain your proof of purchase for warranty claims. The warranty covers defects in materials and workmanship under normal use.
9.2. Onibara Support
For technical assistance, service, or to inquire about calibration services, please visit the official Fluke website or contact their customer support. You can also refer to the comprehensive User Manual (PDF) available online for more detailed information: